Mahr MarSurf PS 10 Traducción Del Manual De Instrucciones Original página 40

Ocultar thumbs Ver también para MarSurf PS 10:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 95
Lc filter
Drive
Measuring speed
Probe type
Calibrate
Parameters
Tolerance limits
Parameter
settings
MOTIF settings
Filter
Arguments
12
Standard
Standard cutoff (as per ISO, JIS, ANSI/ASME).
Short
Next shortest cutoff.
0.8 mm
Fixed cutoff.
Cutoff freely defined.
Lc
mm
On
Drive unit of MarSurf PS 10 is switched on.
Off
Drive unit of MarSurf PS 10 is switched off (e.g. when
the measuring object is moved by a third-party drive).
The third-party drive must be set to the selected
measuring speed.
1.0 mm/s
Measuring speed at which the drive unit moves the
0.5 mm/s
probe over the measuring object.
350 µm
Measuring range of the probe used.
150 µm
100 µm
Calibrates the probe (see Section 3).
Ra
Activated parameters will be evaluated.
Rq
...
Ra 0.000 0.000
Lower and upper tolerance limit for each parameter.
...
Upper section level C1 and lower section level C2 for
C1
µm
evaluating RPc and RSm.
C2
µm
C1 / C2
Vertical counting threshold for evaluating RSm.
Rz / 10
Reference line evaluating the material ratio Rmr or tp.
CREF
%
Section level for three possible material ratio values
C
µm
Rmr or tp.
C
µm
C
µm
Operators A and B for maximum length of roughness
A
mm
motifs / waviness motifs.
B
mm
Upper and lower intersection line for zone width CR
Scr1
%
of the profile peak zone.
Scr2
%
Upper and lower intersection line for zone width CF
Scf1
%
of the profile core zone.
Scf2
%
Upper and lower intersection line for zone width CL
Scl1
%
of the profile valley zone.
Scl2
%
Mahr GmbH, MarSurf PS 10

Publicidad

Tabla de contenido
loading

Tabla de contenido