4. Device setup

Note: For a better understanding of each
step, please fold out pages ΙΙΙ and VΙΙΙ.
4.1 Connecting the device,
performance check
The equipment supplied includes these
items:
base unit, footplate bracket (11), footplate
(7), control pendant (22), patient chip
card (for ARTROMOT-SP3 with chip card
only), power cord (not shown), operating
instructions
1. Set up the device and connect the re-
quired cables by reversing the transport
instructions given in section 6.3.
2. Turn the power switch (16) on.
Caution!
m
Equipment damage –
Be sure to lift the moving bracket for
dorsal extension / plantar flexion (15) a
little, relieving the strain on it, before
you either press the locking pins for the
height adjustment together or loosen
the clamping lever for adjustment of the
inclination (5).
3. Release the locking device (19). To do so,
first relieve the strain on the carriage by
lifting the moving bracket for dorsal ex-
tension / plantar flexion (15) a little. Then
press the two black pins of the locking
device together simultaneously (Fig. A).
4. Set the carriage to an angle of approx.
60°, then release the pins of the locking
device. Check that the pins engage in the
perforation (20).
5. Again, hold the moving bracket for
dorsal extension / plantar flexion (15) and
release the clamping lever for adjust-
ment of the inclination (5). Then lower
the moving bracket for dorsal extension /
plantar flexion (15) until it forms an angle
of 45° with the base. Close the clamping
lever to lock the carriage in that position
(Fig. B).
6. Follow these steps to set the carriage to
the home position:
ARTROMOT®-SP3
without patient chip card
Press the MENU button on the control pen-
dant twice to access programming level 2
(you advance one level each time you press
the button).
Press the "New Patient" parameter button
and select this function (a check
mark appears in the circle next to the
function).
Press the START button. The carriage auto-
matically enters the home position.
ARTROMOT®-SP3
with patient chip card
– Initial adjustment for new patients
Insert the original patient chip card (21)
into the control pendant (22).
Press the MENU button on the control
pendant three times to access program-
ming level 3 (you advance one level each
time you press the button).
Press the "New patient" parameter but-
ton
and select this function (a
check mark appears in the circle next to
the function)
Press the START button.
The CPM device automatically enters the
home position.
– Adjustment with programmed chip card
Insert the original patient chip card (21)
into the control pendant (22).
Press the START button.
The CPM device automatically enters the
home position.
Performance check
If the control pendant can be operated as
described above and the ARTROMOT®-SP3
enters the home position (for home position
52