Ranges & Accuracies:
AC Converter:
61-732 model is averaging sensing, rms calibrated
61-736 model is true rms sensing.
Accuracy:
Accuracy is specified as +/-(a percentage of the reading + a fixed amount) at
23°C±5°C (73.4°F ± 9°F), less than 75% relative humidity.
Temperature Coefficient: 0.1 times the applicable accuracy specification per
degree C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F)
Function
40.0A/400.0A
AC Current*
40.0A/400.0A
400.0V
AC Voltage
600V
DC Voltage 400.0V/600V
400.0Ω/4.000kΩ/40.00kΩ/
400.0k Ω
Resistance
4.000MΩ
40.00MΩ
Audible indication < 25Ω
Continuity
Response time: 500ms
* Accuracy stated for crest factor ≤ 3
Input Impedance is 10MΩ.
Range and
Resolution
(50Hz - 60Hz)
(60Hz - 400Hz)
(50Hz - 500Hz)
(50Hz - 500Hz)
Accuracy
61-732
1.7% + 6 digits 1.7% + 10 digits
3.0% + 6 digits 3.0% + 10 digits
1.2% + 5 digits 1.2% + 8 digits
1.5% + 5 digits 1.5% + 8 digits
0.5% + 2 digits
1.0% + 4 digits
1.5% + 4 digits
3.0% + 5 digits
N/A
Overload
Protection
61-736
400AAC
600 VDC or AC rms
600 VDC or AC rms
600 VDC or AC rms
600 VDC or AC rms
Page 10