HOW TO USE THE MODEL C AIR DERMATOME
(CONTINUED)
9
The desired thickness of the graft is
predetermined by the operator and set by using
the pointer on the thickness guide scale. The
actual thickness of the harvested tissue is highly
dependent upon the operator technique and
the condition of the tissue being harvested. The
thickness guide scale provides both inch and
millimeter values.
Much of the weight of the motor should be
carried by the hand that holds the dermatome
to maintain a reasonably light pressure between
the cutting edge and the patient's body. The
instrument should be held at a 25° - 45° angle.
The cut graft automatically folds itself into the
"pocket" of the dermatome head. To sever the
graft, simply lift the dermatome up and away
from the donor site.
Friction from the rapid back and forth movement
of the blade generates heat, which may cause
a heat related injury to the patient. Once the
dermatome is placed into contact with the
patient you must maintain forward movement
until the graft is taken. Do not allow the
dermatome to remain running for a period
exceeding 30 seconds. If multiple grafts are to
be taken allow a cool down period of not less
than 5 minutes between grafts.