Warning:
The ME device may not be stacked, arranged or used directly next to or with other
devices. When operation is required to be close to or stacked with other devices, the
ME device and the other ME devices must be observed in order to ensure proper
operation within this arrangement. This ME device is intended for use by medical
professionals only. This device may cause radio interference or interfere with the
operation of nearby devices. It may become necessary to take appropriate corrective
measures, such as redirecting or rearranging the ME device or shield.
The rated ME device does not exhibit any basic performance features in the sense of
EN60601-1, which would present an unacceptable risk to patients, operators or third
parties should the power supply fail or malfunction.
Warning:
Portable RF communications equipment (radios) including accessories, such as an-
tenna cables and external antennas should not be used in closer proximity than 30
cm (12 inches) to parts and cables of the ri-magic HPLED specified by the manu-
facturer. Failure to comply may result in a reduction of the device's performance
features.
Guidance  and  manufacture's  declaration  –  electromagnetic  emission
The  Examination  Light  ri-­magic®  HPLED  is  intended  for  use  in  the  electromagnetic  environment  specified  
below.  The  customer  of  the  user  of  the  Examination  Light  ri-­magic  should  assure  that  it  is  used  in  such  an  
environment.
Emission  test
RF  emissions
CISPR  11
RF  emission
CISPR  11
Harmonic  emissions
IEC  61000-­3-­2
Voltage  fluctuations/  
flicker  emissions
IEC  61000-­3-­3
12
Compliance
Electromagnetic  environment  –  guidance
The  Examination  Light  ri-­magic®  HPLED  use  
RF  energy  only  for  its  internal  function.  
Group  1
Therefore,  its  RF  emissions  are  very  low  and  
are  not  likely  to  cause  any  interference  in  
nearby  electronic  equipment.  
The  Examination  Light  ri-­magic®  HPLED  is  
Class  A
suitable  for  use  in  all  establishments,  including  
domestic  establishments  and  those  directly  
connected  to  the  public  low-­voltage  power  
supply  network  that  supplies  buildings  used  for  
Class  A
domestic  purposes.
Complies

Publicidad