Settings - Features
Settings
Thermo Scientific
Features
is used to enable/disable immersion circulator features.
Features
Multi Function Port
Press
next to the desired feature to enable/disable.
Remote Sensor
The
tor requires a 4-pin connector that must mate to a LEMO # ECP.1S.304.
CLL. The immersion circulator uses a 3 wire sensor, but a 4 wire sensor
can be used (pins 3 and 4 are interconnected in the control head). The
pin-out is:
Pin 1 and 2 = Pt100 + Pin 3 and 4 = Pt100 -
Low Level Warning
When
alarm if a low level condition occurs.
Auto Restart
When
was running, after a power interruption condition. If a ramp was running
when power failed, the ramp program resumes where it left off. Note Con-
sider any possible risks before enabling this mode of operation.

Energy Saving

The
ning under a stable load. Enabling the mode saves energy by reducing
the circulator's heater power and cooling requirements. This can result in
substantial energy savings over the life of the circulator. The default set-
ting is disable (enable for A45HC refrigerated circulators).
The refrigeration icon
enabled.
Features
Remote Sensor
X
Low Level Warning
Auto Restart
Energy Saving
X
connector on the rear of the immersion circula-
is enabled, the circulator sounds an
is enabled, the circulator automatically restarts, if it
mode is primarily designed for applications run-
Energy Saving
blinks when the
Section 4 Operation
mode is
4-49