Maintenance; Clean The Instrument; Replace A Fuse; Replacement Parts - Hach IO9000 Instrucciones Para El Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 37

Maintenance

Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.

Clean the instrument

Clean the exterior of the instrument with a moist cloth and a mild soap solution and then wipe the
instrument dry.

Replace a fuse

Electrocution hazard. Remove all power from the instrument and relay connections before this
maintenance task is started.
Fire hazard. Use the same type and current rating to replace fuses.
Item to collect: Fuse, 5 x 20 mm, 1 A, 250 V, fast
Replace a blown fuse for correct operation. Refer to
location.
1. Remove all power to the relay connections.
2. Remove the covers. Refer to
3. Pull the blown fuse out of the fuse holder.
4. Push a new fuse into the fuse holder.
5. Install the covers.

Replacement parts

Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Description
Fuse, 5 x 20mm, 1 A, 250 V, fast
Cable strain relief, 4.3 to 11.4 mm (0.17 to 0.45 in.) cable
Washer, sealing, ½-in. NPT
Nut, Nylon, Jam, ½-14, NPT
18 English
D A N G E R
D A N G E R
D A N G E R
Wiring overview
Remove the covers
on page 10.
W A R N I N G
on page 10 for the fuse
Item no.
180003201
5202500
1033814
8774

Publicidad

Tabla de contenido
loading

Tabla de contenido