MP6025871-12C1 / A
Guidance and manufacturer's declaration
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The Guardian 2 Link is intended for use in the electromagnetic environment specified
below. The customer or the user of the Guardian 2 Link should make sure that it is
used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions
CISPR 11
RF emissions
CISPR 11
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Guardian 2 Link is intended for use in the electromagnetic environment specified
below. The customer or the user of the Guardian 2 Link should assure that it is used in
such an environment.
Immunity Test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short in-
terruptions and voltage
variations on power
supply lines
IEC 61000-4-11
Group 1
The Guardian 2 Link must emit electromagnetic energy in order to
perform its intended function. Nearby electronic equipment may
be affected.
Class B
The Guardian 2 Link is suitable for use in all establishments, in-
cluding domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used for
domestic purposes.
IEC 60601 Test Level
±2 kV, ±4 kV, ±8 kV Air
±2 kV, ±4 kV, ±6 kV Indirect
±2 kV for power supply
lines
±1 kV for input/output lines
±1 kV line(s) to line(s)
±2 kV line(s) to earth
<5% U
(>95% dip in U
T
for 0.5 cycle
Compliance
Level
±8 kV Air
±6 kV Indirect
±22 kV Air,
<5% RH
Not applicable
Not applicable
)
Not applicable
T
-16-
Electromagnetic Envi-
ronment - Guidance
For use in a typical do-
mestic, commercial, or
hospital environment.
Requirement does not ap-
ply to this battery pow-
ered device.
Requirement does not ap-
ply to this battery pow-
ered device.
Requirement does not ap-
ply to this battery pow-
ered device.