4. Safety instructions
4.1 Safety instructions
EN
14
Safety instructions
For your own safety and in accordance with the require-
ments of the 93/42/EEC, we would draw your attention to
the following points.
General
Read this instruction manual carefully. It forms
part of the device and must be available at all
times.
Warning!
•
Do not use the MEDUCORE Easy in an oxygen-en-
riched atmosphere or in the presence of flamm- able
anaesthetics or other substances/air mixtures at risk of
fire. If this instruction is not followed, there is a risk of
explosion.
•
Use the device only in dry places which do not con-
duct electricity. If this instruction is not followed, inju-
ries (electric shock, burns) may be sustained by the
patient, the user or bystanders.
•
Never use the device on patients with a body weight
of less than 20 kg. If this instruction is not followed,
the patient may sustain life-threatening injuries.
•
Before using the MEDUCORE Easy, examine the pa-
tient who is suspected of an acute cardiovascular ar-
rest for vital signs such as responsiveness and
breathing. MEDUCORE Easy must only be used to de-
fibrilate non-responsive patients who are not breath-
ing properly.
•
Before using the MEDUCORE Easy, ensure that nei-
ther the device nor its accessories are visibly damaged.
If you should find damage on the device or the acces-
sories, do not use it. If this instruction is not followed,
malfunctions may occur causing injuries to users and
patients.