4 User interface and navigation
Section 1 Specifications
Specifications are subject to change without notice.
Specification
Dimensions (W x H x D)
Enclosure
Weight
Installation category
Pollution degree
Protection class
Altitude
Display
Mounting
Power requirements
Power connection
Operating temperature
Storage temperature
Humidity
Air purge (optional)
Alarm relay outputs
Alarm connection
Recorder output
Recorder connection
Maximum reagent usage 0.5 L per month (for each of the two standards and one reagent)
Reagent containers
Measurement range
Accuracy/precision
Cycle limit
Calibration
Detection limit
Certifications
on page 4
on page 16
Details
34.11 x 47.50 x 17.93 cm (13.43 x 18.70 x 7.06 in.)
IP62 rating when the door is latched (indoor use)
11.3 kg (25 lb)
II
2
I
2000 m (6560 ft)
LCD, 3.5-digit measurement readout and six-character alphanumeric scrolling text line
Wall mount
100-115/230 VAC; 90 VA, 50/60 Hz; 2.5 A fuse
12 to 18 AWG wire
5 to 40 °C (41 to 104 °F)
-40 to 60 °C (-40 to 140 °F)
90% at 40 °C (90% at 104 °F)
0.1 CFM instrument-quality air, ¼-inch O.D. tubing
Two unpowered SPDT relays, each rated at 5 A resistive, 240 VAC maximum.
12 to 18 AWG wire
One isolated recorder output, 4-20 mA (can be adjusted to 0-20 mA), maximum
impedance 500 ohms
12 to 18 AWG wire
High-density polyethylene 0.5 L bottles contained inside the analyzer enclosure and
vented externally.
0.1 to 10 mg/L fluoride
±10% or ±0.10 ppm; ±7% or ±0.07 ppm
4.2 minutes
Two-point log/linear with 0.5 mg/L and 5.0 mg/L fluoride standards
0.10 ppm
CE. Listed to UL and CSA safety standards by ETL.
English 3