D D i i s s i i n n f f e e c c t t i i n n g g
W W A A R R N N I I N N G G
• Always disinfect the support surface between patients to avoid the risk of cross-contamination and infection.
• Always make sure that you wipe each product with clean water and dry each product after you clean. Some cleaning
agents are corrosive in nature and may cause damage to the product. Failure to follow these cleaning instructions may
void your warranty.
• Do not immerse the support surface in cleaning or disinfectant solutions.
• Do not allow fluid to pool on the support surface.
C C A A U U T T I I O O N N
• Always dry the support surface covers before you store, add linens, or place a patient on the surface. Dry the product to
help prevent the performance of the product from impairment.
• Do not over expose the covers to higher concentration disinfectant solutions as these may degrade the covers.
• Do not allow liquid to seep into the zipper area or watershed cover barrier when you clean the underside of the support
surface. Fluids allowed to come in contact with the zipper may leak into the support surface.
• Do not use hydrogen peroxides or quaternaries that contain glycol ethers as they may damage the cover.
Recommended disinfectants:
Quaternaries
Phenolic disinfectants
Chlorinated bleach solution (5.25% bleach 1:100 dilution with water)
70% Isopropyl alcohol
To disinfect the support surface covers after each patient use, follow these steps in order:
1. Clean and dry the support surface covers before you apply disinfectants.
2. Apply recommended disinfectant solution by spray or pre-soaked wipes (do not soak the support surface).
N N o o t t e e - - Make sure that you follow the disinfectant's instructions for appropriate contact time and rinsing requirements.
3. Wipe the support surface covers with a clean, dry cloth to remove any excess liquid or disinfectant.
4. Allow the support surface covers to dry before returning to service.
EN
10
2850-109-005 Rev A.0