3. Unzip the cover. Start at the foot end patient right corner of the support surface and stop at the head end patient right
corner.
4. Fold the top of the cover to the patient's right side and then remove the foam crib assembly from the bed and set aside.
5. Remove and discard the cover.
6. Place the replacement cover, unzipped and open, on the bed with the black bottom cover on the litter and the top cover
folded over the patient's right side of the bed.
7. Place the foam crib assembly on top of the bottom part of the cover to make sure that the foam crib is aligned with the
cover.
8. Fold the top cover over the top of the foam crib assembly to make sure that the top cover is aligned with the foam crib
assembly.
9. Zip the cover to close. Start at the head end patient right corner and stop at the foot end patient right corner.
10. Verify proper operation before you return the product to service.
F F i i r r e e b b a a r r r r i i e e r r r r e e p p l l a a c c e e m m e e n n t t
T T o o o o l l s s R R e e q q u u i i r r e e d d : :
None
P P r r o o c c e e d d u u r r e e : :
1. Raise the bed height to the full up position.
2. Lower the fowler and gatch sections to the full down positions.
3. Unzip the cover. Start at the foot end patient right corner of the support surface and stop at the head end patient right
corner.
4. Fold the top of the cover to the patient's right side.
5. Starting at the foot end, roll up the fire barrier on the foam crib assembly.
N N o o t t e e - - Work from side to side, a little at a time, to move the fire barrier to the top of the support surface.
6. Remove and discard the old fire barrier.
7. Starting at the head end, roll the new fire barrier down and slide over the foam crib assembly.
N N o o t t e e - - Align the fire barrier on the foam crib before sliding over the foam crib assembly.
8. Slide the fire barrier down the foam crib assembly, work from side to side. Make sure that the fire barrier is tight on the
foam crib assembly.
9. Align the foam crib assembly on top of the bottom part of the cover.
N N o o t t e e - - Spread the excess fire barrier material below the foam crib assembly at the foot end.
10. Fold and align the top cover over the top of the foam crib assembly.
11. Zip the cover to close. Start at the head end patient right corner and stop at the foot end patient right corner.
12. Verify proper operation before you return the product to service.