3. Install the Cassette into the Console and ensure the
assembly is locked in place. The Release Lever on the
Cassette will click when installed properly. Refer to
Figure 2.
Release
Lever
Figure 2. Install the Cassette
4. Connect the Cassette Tubing to the Vacuum Canister. Do
not twist tubing during installation. Verify the connection is
secure to prevent a decrease in vacuum. Refer to Figure 3.
Figure 3. Connect Cassette Tubing to Vacuum Canister
5. For Vacuum and Rinse Tubing Cassette only, connect
the Saline Spike to the Saline. Place the Saline in the
Saline Cavity of the Console. Verify the saline tubing is not
kinked. Refer to Figure 4A and Figure 4B.
Figure 4A.
Connect the Saline Spike
6. When ready to begin the biopsy procedure, connect the
Probe Vacuum Tubing to the Cassette Vacuum Tubing
connector. If using rinse, connect the Probe Rinse Tubing
to the Cassette Rinse Tubing connector. Verify the
connections are secure to prevent leakage. Refer to
Figure 5.
Rinse Tubing
Check Valve
Vacuum
Connector
Figure 5. Connect Probe Tubing to Cassette Connectors
7. At the conclusion of the procedure, disconnect all tubing,
remove the Cassette, and dispose of properly.
8. Refer to the E
C
E
N
OR
NSPIRE
Instructions for Use for information on cleaning and
maintenance of the E
Biopsy System.
Warranty
of this product that this product will be free from defects in
materials and workmanship for a period of one year from
product warranty will be limited to repair or replacement of
the defective product, in Bard Peripheral Vascular's sole
discretion or refunding your net price paid. Wear and tear
from normal use or defects resulting from misuse of this
product are not covered by this limited warranty.
Figure 4B.
Install the Saline in the
Saline Cavity
Breast Biopsy System
®
C
E
Breast
®
N
OR
NSPIRE
OR IMPLIED, INCLUDING, BUT NOT LIMITED TO,
OR FITNESS FOR A PARTICULAR PURPOSE. IN NO
LIABLE TO YOU FOR ANY INDIRECT, INCIDENTAL OR
CONSEQUENTIAL DAMAGES RESULTING FROM YOUR
Some states/countries do not allow an exclusion of implied
warranties, incidental or consequential damages. You may be
entitled to additional remedies under the laws of your state/
country.
Extended warranty service agreements are available. Consult a
Bard representative for details regarding terms and conditions.
An issue or revision date and a revision number for these
instructions are included for the user's information on the last
page of this booklet. In the event 36 months have elapsed
between this date and product use, the user should contact
Bard Peripheral Vascular, Inc. to see if additional product
information is available.
Assembled in Thailand
2