Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to maintain a device in accordance
with the instructions may invalidate the device's warranty. Furthermore, failure to maintain a device may compromise the clinical
condition or safety of users and/or their caregivers. Do not perform service and maintenance while the user is in the product.
Contact local dealer for assistance, if needed, in setting up, using or maintenance of the device.
1st Day
EN
Read the User Guide carefully before use and store it in
a safe place for future reference
Daily
Use a dry cloth to clean the product
Visual inspection. Repair or replace any damaged or
worn-out parts
Ensure that all fasteners are in place and fitted
correctly
Be aware of any MAX mark indications before
adjustment
Weekly
Check that all wheels are moving freely and that all
wheel locks/brakes are easily activated
Wipe off any residue and dirt from the product using a
cloth with warm water and a mild detergent/soap and
dry before use
Monthly
All nuts and bolts on this product must be checked and
tightened to avoid unnecessary failures
Oil the swing-away parts. We recommend you use a
professional lubricant system
Annually
Inspect the frame for cracks or signs for worn-out parts
and carry out annual service. Never use a product
which appears to be substandard or faulty
Warewashing machine
Remove the following items and clean them separately:
Cushions, Seat and Accessories
This product can be washed at 60° using a mild detergent
for 10 minutes in a washing machine designed for medical
devices. Use the machine's drying function to dry the product.
Disinfection
The product can be disinfected with a 70% disinfectant IPA
solution. It is recommended to wipe off any residue and
dirt from the product, using a cloth with warm water and a
mild detergent/soap without chlorine and let it dry before
disinfection.
Service interval
A detailed inspection of the product must be made every 12
months (every 6 months if the product is used very often
- daily use for more than 7 hours), every time the product
is reissued for use and after a long term storage (more
than 4 months). The inspection should be carried out by a
technically competent person who has been trained in the
use of the product.
The lifetime of this product in normal use is 8 years if all
maintenance and servicing is carried out in accordance with
the manufacturer's instructions and demonstrably recorded.
The device may be reissued within its lifetime
Service Check List:
Check and adjust the following functions as per their
intended use. Repair or replace any damaged or worn-out
parts. Please check the:
Frame is adjustable in height
Seat plane adjustment. The seat plane is adjustable
and stays in position after adjustment.
Wheels are moving freely and that all wheel locks/
brakes are easily activated. Adjust if required
Nuts and bolts on the product are tightened
Oil the swing-away parts
Product for cracks or signs for worn-out parts
All labels on the product are intact
Materials
The following surface treatment processes were used to
protect against corrosion:
Lacquered surfaces = Polyester powder coating or
ED-coating
Non-lacquered aluminium parts = Anodized coating
Non-lacquered steel surfaces = Galvanized
34