3 Dimensional Drawings

ALLOW MINIMUM CLEARANCE OF
ALLOW MINIMUM CLEARANCE OF
18 in. [457]
18 in. [457]
FOR UV LAMP REPLACEMENT
FOR UV LAMP REPLACEMENT
DO NOT
OPEN WHEN
UV LAMP
IS ON
UV LAMP
REPLACEMENT COVER
AC
HIGH VOLTAGE
POWER
DISCONNECT
POWER FOR
SERVICING
TRANSMITTER
CONNECTION

4 4000TOCe Sensor Installation

Installation
Mount the 4000TOCe in an appropriate location. A minimum clearance dimension of
18" (45.7 cm) is required on the left side of the sensor for UV lamp change-out. Two
plumbing connections are provided on the right side of the enclosure. The lower connec-
tion is labeled 'Sample Inlet' and the upper connection 'Sample Outlet'. A third connec-
tion, a safety drain, is located on the bottom of the enclosure. Figure 1 shows a typical
installed configuration. The installation kit provided includes installation tube assem-
blies, a sample conditioning coil and a high capacity inlet filter. Attach the high capacity
inlet filter to the sample inlet connection as shown in Figure 2.
Figure 1
6
10.05
[255.3]
7.40
7.40
[188]
[188]
Ø0.25
Ø0.25
TYP
TYP
[Ø6.35]
[Ø6.35]
[279.4]
5000TOCi Sensor Dimension Diagram
DIMENSIONS IN INCHES [mm]
Sample
Safety Drain
Atmospheric
Drain
15.0
15.0
[368.3]
REF
[368.3]
10.05
[255.3]
11.0
11
[279.4]
Process Pipe
6 to 36 in.
Drain
[15 to 90 cm]
DIMENSIONS IN
INCHES [mm]
8.05
[204.5]
8.05
[204.5]
DRAIN BEFORE
8.80
SHIPPING
8.8
[223.5]
DO NOT
SAMPLE
[223.5]
OPEN WHEN
OUTLET
UV LAMP
IS ON
SAMPLE
INLET
5.5
5.5
[139.7]
[139.7]
Process Pipe
Cross-Section
Isolation Valve
Regulator
(Optional)
30°-60°
Recommended
10-32
INTERNAL THD
3.26
3.26
[82.8]
2.08
[82.8]
2.08
[52.8]
[52.8]
REGULATOR
REGULATOR
NOT SHOWN
NOT SHOWN
1.64
1.64
[41.6]
[41.6]