Dimensional Drawings; 6000Toci Sensor Installation And Start-Up - Mettler Toledo 6000TOCi Guia De Inicio Rapido

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 59

3 Dimensional Drawings

ALLOW MINIMUM CLEARANCE OF 18
in. [457] FOR UV LAMP REPLACEMENT
CLEARANCE FOR REMOVAL OF UV LAMP

4 6000TOCi Sensor Installation and Start-up

Installation
Mount the 6000TOCi in an appropriate location. A minimum clearance dimension of
18"(45.7 cm) is required on the left side of the sensor for UV lamp change-out. 10"
clearance above the sensor is required for CAL/SST. Two plumbing connections are
provided on the right side of the enclosure labeled 'Sample Inlet' and 'Sample Outlet'.
6
0.6
[14.5]
CLEARANCE FOR INSTALLATION/REMOVAL OF CAL/SST SERVICE TOOL
USED FOR PERIODIC CALIBRATION, VERIFICATION AND VALIDATION
12.50 [ 318 ]
12.8
[325.7]
11.9
[302.75]
30.5 [ 775 ]
4X Ø 0.260 [6.600] THRU ALL
INSTALL WITH M6 SIZE SCREW
WALL BRACKET
8.56 [ 217 ]
11.92 [ 303 ]
TOC/SST OVERALL WIDTH
ALLOW MINIMUM CLEARANCE OF 10
in. [254]
9.0
[229.8]
5.7 [144.7]
CAL/SST MODULE
6000TOCi SENSOR
6.10 [ 155 ]
1.64
[ 42 ]
DIMENSIONS IN
INCHES [mm]
5.4
[137.2]
0.8 [21.53]
2.2 [55.07]

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

M800

Tabla de contenido