12 Specifications continued
Installation/Power/Enclosure
Inlet Connection
Outlet Connection
Inlet Filter
Wetted Parts
Wall Mount
Maximum Sensor Distance
* Readout in equivalent S/m ranges selectable at M800
** Temperature above 70°C requires Sample Conditioning Coil (included)
*** For power plant cycle chemistry samples, pH may be adjusted by measurement after cation exchange.
**** Process pressure above 85 psig (5.9 bar(g)) requires optional High Pressure Regulator p/n 58 091 552.
Specifications subject to change without notice.

13 Service and Maintenance

The 6000TOCi Sensor is designed to minimize service and maintenance. There are
virtually no moving mechanical components, therefore normal operating wear and tear is
minimal. This reduces the amount of consumable components as well as the amount of
time needed to maintain the sensor. Listed below are instructions on how to perform
simple periodic maintenance, which includes UV Lamp change (every 4500 hours of
operation), filter replacement (typically every 6 months), and general cleaning.
UV Lamp Replacement
WARNING: UV RADIATION HAZARD
Apply power to UV lamp only when installed in housing in accordance with
operation manual. DO NOT remove UV lamp from housing unless power is off.
Always protect eyes and skin from direct exposure to UV light.
METTLER TOLEDO Thornton recommends replacement of the UV lamp inside the 6000TOCi
Sensor after 4500 hours of operation, or 6 months of continuous use, not to exceed one year.
This is a simple procedure that requires only a few minutes to complete. The following steps
explain the proper procedure for the UV lamp replacement. Refer to the illustration below.
0.125" [3 mm] O.D. (6' [2 m] FDA compliant PTFE tubing supplied)
0.125" [3 mm] O.D. (6.5" [165 mm] fixed 316 SS tube provided)
316 SS, inline 60 micron
316 SS / Quartz / PEEK / Titanium / PTFE / EPDM
Standard, mounting bracket provided
300 ft [91 m]
Lamp
13