5 Electrical Connection

WARNING: MAIN POWER MUST EMPLOY A SWITCH OR CIRCUIT BREAKER
AS THE DISCONNECTING DEVICE FOR THE EQUIPMENT. THE BREAKER
SHOULD BE PROPERLY LABELED AND EASILY ACCESSIBLE FOR SYSTEM
ISOLATION BY SERVICE PERSONNEL. THE CIRCUIT BREAKER SHALL HAVE A
MAXIMUM OF 20 AMP RATING.
Two fittings are provided on the left side of the enclosure: the AC Power bulkhead fitting
for power cable, and the M800 patch cable connection. Feed the power cable through
the bulkhead fitting on the side-wall. Remove the two thumb screws protecting the TOC
circuit board. Line and Neutral AC power connections are positioned on the top-left
side of the TOC enclosure. The AC Ground connection is provided on a grounding plate
located to the left of the Line and Neutral terminals. See Figure 3 for terminal wiring. See
product label for specified fuse size.
Figure 3
LINE
LINE
EARTH GROUND
EARTH GROUND
NEUTRAL
NEUTRAL
6 Terminal Block (TB) definitions M800 Transmitter
6000TOCi Sensor Cable Connection
TOC Sensor uses 58 080 27X cables.
TB2 and TB4 – Terminal Assignment for 6000TOCi
TB2
(ISM Ch1,2)
Term. Function
1
DI2+
2
DI2–
3
1-Wire_Ch1
4
GND5V_Ch1
5
RS485B_Ch1
6
RS485A_Ch1
7
GND5V_Ch1
8
5V_Ch1
9
24V_Ch2
10
GND24V_Ch2
11
1-Wire_Ch2
12
GND5V_Ch2
13
RS485B_Ch2
14
RS485A_Ch2
15
GND5V_Ch2
16
5V_Ch2
*For M800DP Transmitters equipped with Profibus DP communication capabilities, please refer to the M800DP Trans-
mitter Operation Manual.
8
TB4
6000TOCi
(ISM Ch3,4)
Function
Sensor wire color
DI6+
DI6–
1-Wire_Ch3
GND5V_Ch3
RS485B_Ch3
black
RS485A_Ch3
red
GND5V_Ch3
white
5V_Ch3
blue
24V_Ch4
GND24V_Ch4
1-Wire_Ch4
GND5V_Ch4
RS485B_Ch4
black
RS485A_Ch4
red
GND5V_Ch4
white
5V_Ch4
blue