12 Conductivity Quick Setup
Sensor Type Selection
1.25
µS/cm
25.0
°C
Sensor Type = Cond(2)
Cell Constant
1.25
µS/cm
25.0
°C
p M=0.1003
A=0.0000
s M=1.0000
A=0.0000
Measurement units
1.25
µS/cm
25.0
°C
a S/cm
13 Conductivity Calibration for 2-e and 4-e Sensors
Cond One point Calibration
1.25
A
µS/cm
25.0
A
°C
Calibrate Sensor
Channel A Conductivity
1.25
H
µS/cm
25.0
A
°C
Cal Compensation
Standard
(PATH: Menu / Quick Setup)
Select the type of sensor to be used with the M400 transmitter. Choices are "Cond(2)",
used for all 2-electrode sensors and "Cond(4)" for all 4-electrode sensors. Press the
[ENTER] key.
u
Enter the appropriate cell constant(s):from the sensor label or certificate (M) for
2-electrode sensors, leaving (A) at 0.000; or (M) and (A) values for 4-electrode
sensors. Press the [ENTER] key.
Select the measurement (conductivity or temperature) and units for measurement.
While in Measurement mode press the [CAL] key. If the display prompts you to enter
the calibration security code, press the m or . key to set the calibration security
code, then press the [ENTER] key to confirm the calibration security code.
u
Press [ENTER] to select the CONDUCTIVITY CALIBRATION SUB FUNCTION.
A flashing "H" shows the ongoing calibration process.
Choose the compensation mode by using the m or . key.
Press [ENTER] to confirm the compensation mode.
u
(PATH: Cal)
23