4
Enter the value for the conductivity which belongs to the second concentration
at the first temperature and press [ENTER] etc.
After entering all conductivity values, that belong to the different concentra-
tions at the first temperature point, enter in the same way the value of the 2nd
temperature point (Temp2) and the conductivity value which belongs at the
second temperature to the first concentration. Press [ENTER] and go on in the
same way for the next concentration points as described for the first tempera-
ture point.
Enter in this way the values at every temperature point. After entering the last
value, pressing [ENTER] again will bring up the Save Changes dialog. Select-
ing No will discard the entered values and return to the measurement display
screen, selecting Yes will save changes made.
NOTE: The values for the temperature have to increase from Temp1
h
to Temp2 to Temp3 etc. The values for the concentration have to
i ncrease from Concentration1 to Concentration2 to Concentration3
etc.
NOTE: The conductivity values at the different temperatures have
h
to increase or decrease from Concentration1 to Concentration2 to
Concentration3 etc. Maxima and/or minima are not permitted. If the
© Mettler-Toledo AG, CH-8606 Greifensee
M400 Transmitter: New Functions
Enter the values for the different concentra-
tions (ConcentrationX).
Press [ENTER].
Enter the value of the 1st temperature
(Temp1) and the value for the conductivity
which belongs to the first concentration at this
temperature.
Press [ENTER].
Printed in Switzerland