M400 Transmitter: New Functions
conductivity values at Temp1 are increasing with the different concen-
trations, they have to increase also at the other temperatures. If the
conductivity values at Temp1 are decreasing with the different con-
centrations, they have to decrease also at the other temperatures.
2
DLI, TTM, ACT Assignment
to Analog Outputs and Relays
NOTE: Beside the measurement values pH, O2, T, etc. also the ISM
h
values DLI, TTM and ACT can be assigned to the different lines in the
display (see chapter 8.2.1.2 "ISM sensors") and linked to the analog
outputs (see chapter 8.3 "Analog outputs") or set points (see chapter
8.4 «Set points").
3
Analog Input Signal for Pressure Compensation
TB3 – 4 to 20 mA input signal
Pin no.
1
2
3
4
5
6
7
8
9
NOTE: Jumper 3 to 4 has to be installed. 50 Ohm resistor has to be
h
installed between 1 and 4.
© Mettler-Toledo AG, CH-8606 Greifensee
Function
! input of 4/20 mA signal
not used
not used
@ input of 4/20 mA signal
not used
not used
not used
not used
not used
5
Printed in Switzerland
30 015 281