M400 Transmitter New Functions
4
Temperature Configuration for CIP / SIP Counter
CIP Cycle Limit
If the Max setting is on 000, the counter functionality is turned off. The limit
for CIP recognition will be configured with the parameter Temp.
SIP Cycle Limit
If the Max setting is on 000, the counter functionality is turned off. The limit
for SIP recognition will be configured with the parameter Temp.
© Mettler-Toledo AG, CH-8606 Greifensee
The CIP cycle limit counts the number of CIP
cycles. If the limit (user defined) is reached,
an alarm can be indicated and set to a certain
output relays. The following options are pos-
sible:
CIP Max 000
Alarm
YES / NO
The SIP cycle limit counts the number of SIP
cycles. If the limit (user defined) is reached,
an alarm can be indicated and set to a certain
output relays. The following options are pos-
sible:
SIP Max 000
Alarm
YES / NO
Temp 055
R#
choose relay
Temp 115
R#
choose relay