Manual RRC-SMB-FPC
RRC SMBUS FLATPAQ C
HARGER
Manual RRC-SMB-FPC
RRC SMBUS FLATPAQ C
HARGER
.en
This battery charger may not be located in the vicinity of a patient.
In accordance with EN60601-1-1, the horizontal distance must be at least
1.5 meters.
Do not insert a fully charged battery into the charger repeatedly. Due to the
high initial charge, the battery could be overcharged which could lead to
damage to the battery and to the battery charger
Use the battery charger only in dry indoor environments! Wet and moisture
covered products should not be used!
This product has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate
FCC Advice,
radio frequency energy and, if not installed and used in accordance with the
valid in USA
instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to
radio or television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
•
Reorient or relocate the receiving antenna.
•
Increase the separation between the equipment and receiver.
•
Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
•
Consult the dealer or an experienced radio/TV technician for help.
Changes or modifications of the device are not allowed.
Declaration of
The battery charger complies to the current regulations of the EU guidelines.
Conformity:
Symbols
Caution / Warning
Instructions for use
Use only in enclosed spaces
Sign of conformity to the applicable EC Directives
The product should be disposed of according to local guidelines
RoHs (China)
DC voltage
Technical
Battery Charger
Specifications
Input Voltage
19 - 26VDC / 2.8A max.
Charging and
0 - 17.4VDC / 0 - 4A
Output Voltages
Power
40W max.
Measurements
88.5 x 38 x 20 mm
(L x B x H)
Operating/Storage
0°C to 40°C / -10°C to 70°C
Temperature
Weight
36g
Safety Standards /
EN60950-1* / IEC/UL60950-1* / EN55011 / EN55032 / EN55024* / FCC15 class B* /
EMC Standards
EN/IEC61000-4-2,-3,-4,-5,-6,-8,-11* / EN61000-3-2,-3-3*
* valid only for power supply
E
3
NGLISH
Power Supply
100-240VAC / 50 - 60Hz
19VDC ± 5% / 3.4A
65W
95 x 51 x 25.4 mm
0°C to 40°C / -10°C to 70°C
270g