Electromagnetic immunity

The AIR-FLOW MASTER is intended for use in the electromagnetic environment specified below. The customer or the user of the AIR-FLOW MASTER
should assure that it is used in such an environment.
immunity test
iEC 60601 test level
Electrostatic discharge
± 6 kV contact
(ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast transient
± 2 kV for power supply lines
/ burst
IEC 61000-4-4
Surge
± 1 kV differential mode
IEC 61000-4-5
± 2 kV common mode
Voltage dips, short in-
<5 % U
terruptions and voltage
(>95 % dip in U
variations on power
40 % U
supply input lines
(60 % dip in U
IEC 61000-4-11
70 % U
(30 % dip in U
<5 % U
(>95 % dip in U
Power frequency
3 A/m
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE:
U
is the a.c. main voltage prior to application of the test level.
T
FB-348_3_ed_2008-06_EMC AIR-FLOW MASTER.indd 5
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV differential mode
± 2 kV common mode
<5 % U
T
) for 0.5 cycle
(>95 % dip in U
T
40 % U
T
) for 5 cycles
(60 % dip in U
T
70 % U
T
) for 25 cycles
(30 % dip in U
T
<5 % U
T
) for 5 sec
(>95 % dip in U
T
3 A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic mate-
rial, the relative humidity should be at least
30 %.
Main power quality should be that of a typical
commercial or hospital environment.
Main power quality should be that of a typical
commercial or hospital environment.
Main power quality should be that of a typi-
T
) for 0.5 cycle
cal commercial or hospital environment. If
T
the user of the AIR-FLOW MASTER requires
T
) for 5 cycles
continued operation during power main inter-
T
ruptions, it is recommended that the AIR-
T
) for 25 cycles
FLOW MASTER be powered from an uninter-
T
ruptible power supply or a battery.
T
) for 5 sec
T
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
5 5