Electromagnetic immunity, rF portable equipment
The AIR-FLOW MASTER is intended for use in the electromagnetic environment specified below. The customer or the user of the AIR-FLOW MASTER should
assure that it is used in such an environment.
immunity test
iEC 60601 test level
Portable and mobile RF communications equipment should be used no closer to any part of the AIR-FLOW MASTER, including cables, than the recom-
mended separation distance calculated from the equation applicable to the frequency of the transmitter.
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey
range
.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
6 6
FB-348_3_ed_2008-06_EMC AIR-FLOW MASTER.indd 6
Compliance level
Electromagnetic environment - guidance
Recommended separation distance
3 Vrms
d = 1.2 √P
3 V/m
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in metres (m).
AIR-FLOW MASTER, including cables, than the recom-
, including cables, than the recom-
, should be less than the compliance level in each frequency
a
02/06/08 12:50:32