Assembly and Replacement of
the Tunable Diaphragm
Using your fingertip, apply a small
amount of talcum powder to the
inside surface of the flexible edge
of the diaphragm. This will
facilitate assembly and maintain
smooth and quiet low/high
frequency alternation. Snap the
diaphragm with your fingers to
remove excess talc.
Insert the flexible edge of the
diaphragm into the groove of the
rim. This is best accomplished
by starting with the rim
positioned above the legible
side of the diaphragm.
Visually examine the ring to
ensure that the flexible edge is
smoothly engaged inside the rim.
If necessary, flex the assembly
by pinching the outer rim edge
between the thumb and fingers.
Repeat this flexing procedure
after rotating the assembly
one quarter.
To attach the rim/diaphragm
assembly to the chestpiece,
engage the groove of the rim
(with the diaphragm attached)
around the chestpiece at one
point and hold it in place with
your thumbs.
Slowly roll the rim around and
over the chestpiece edge using
both thumbs, moving in opposite
directions around the chestpiece.
Visually inspect the edge where
the diaphragm engages the rim
for uniform containment. Minor
adjustments may be made by
slightly pulling and rolling the rim
away from the diaphragm,
allowing the diaphragm to slip
into position.
Special Procedures Adaptor
With the 3M
Cardiology
complete seal of the chestpiece
against the skin is desirable. The
Special Procedures Adaptor
ensures a complete seal for small
site auscultation by reducing the diameter of the
chestpiece. It is flexible and conformable and
especially useful for pediatric or thin patients, around
bandages and carotid assessment.
The Special Procedures Adaptor is designed to
fit over the diaphragm side of the chestpiece
and enhances both low and high frequency
detection by improving source localization.
Attach the adaptor to the chestpiece by pressing
it over the diaphragm. It may be easily removed
by pulling it away from the chestpiece. It may be
wiped clean with alcohol or warm soapy water.