Diagnostic Hardware
Assembling the Cysto Sheath and Obturator
1.
Insert the obturator into the cysto sheath, aligning the "0" on
the obturator (a) with the "0" on the cysto sheath (b) (see
Figure 4).
F F i i g g u u r r e e 4 4 . . I I n n s s e e r r t t i i n n g g t t h h e e o o b b t t u u r r a a t t o o r r i i n n t t o o t t h h e e c c y y s s t t o o s s h h e e a a t t h h
2.
Twist the obturator to lock it in place.
3.
Untwist the obturator to remove it from the cysto sheath.
Assembling the Cysto Sheath, Bridge, and Scope
1.
Align the "0" on the bridge (a) with the "0" on the cysto sheath
(b) and insert the bridge into the cysto sheath (see Figure 5).
F F i i g g u u r r e e 5 5 . . I I n n s s e e r r t t i i n n g g t t h h e e b b r r i i d d g g e e i i n n t t o o t t h h e e c c y y s s t t o o s s h h e e a a t t h h
2.
Twist the bridge to lock it into place.
3.
Slide the scope into the bridge speed-lock mechanism (a)
until the two pieces snap together (see the Figure 6).
F F i i g g u u r r e e 6 6 . . I I n n s s e e r r t t i i n n g g t t h h e e s s c c o o p p e e i i n n t t o o t t h h e e b b r r i i d d g g e e s s p p e e e e d d - - l l o o c c k k
4.
Press the speed-lock button on the bridge to remove the
scope.
5.
Untwist the bridge to remove it from the cysto sheath.
EN- 10