Instructions For The Calibration Check; Contact Information For Customer Queries - Hartmann Tensoval duo control Instrucciones De Uso

Ocultar thumbs Ver también para Tensoval duo control:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

by the legislator, such as, in
Germany, the "Medizinprodukte-
Betreiberverordnung" (Medical
Device Operating Regulation).
15.2 Instructions for the
calibration check
The calibration check can be carried
out by the competent authorities or
authorised maintenance providers
against compensation. A device
function check can be carried out
on people or using a suitable simu-
lator. Calibration check involves
testing for leak tightness of the
pressure system and possible devia-
tions of the pressure reading.
Remove at least one battery in
order to switch to calibration mode.
Hold down on the START / STOP
button and then insert the battery.
Release the button and, after a few
moments, two zeros will appear
one above each other on the dis-
play. Instructions on the calibration
check will be supplied on request
to competent authorities or
authorised maintenance providers
by HARTMANN.
15.3 Contact information for
customer queries
AE – PAUL HARTMANN
Middle East FZE
Dubai
AU– PAUL HARTMANN Pty. Ltd.
Level 6, 5 Ryder Boulevard
Rhodes, NSW 2138
Australia
HK – PAUL HARTMANN
Asia-Pacific Ltd.
Hong Kong
ZA – HARTMANN South Africa
2194 Johannesburg
Date of revision of the text:
2011-07
118

Publicidad

Tabla de contenido
loading

Tabla de contenido