Etac Molift Smart 150 Manual De Usuario página 11

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 116
Transport and operating conditions
Transport and storage
For long time storage it is recommended that the
emergency stop button is activated (pushed in).
The lifter can be stored and transported under
temperatures between -25 - 70 °C.
Operating
The lifter is designed for use at standard room
temperatures (+5 to +40°C).
Air pressure: 70 - 106 kPa
Relative humidity: 15 - 93 %
Following storage or transport at other tempera-
tures leave the lifter in a room with a suitable
temperature until it reaches a safe operating
temperature.
Medical electrical equipment
requires special precautions regard-
ing electromagnetic compatibility
(EMC). Portable or mobile radio
communication equipment may
affect the medical electrical equip-
ment, and should be kept minimum
25 cm (10 inches) from the lifters
electronics.
Battery
Molift Smart 150 is equipped with a 14.4 V 2.6Ah
NiMH battery. The battery has a life expectancy of
approx. 500 charging cycles.
Recharge battery upon receipt, and
then every six months, to avoid
battery failure.
Used batteries must be disposed
of as special waste according to
local rules and regulations. Do not
dispose in general waste
Batteries must be fully charged
and drained completely a couple of
times before it achieves full capac-
x2
ity. Also follow this procedure if
the lifter has not been in use for a
long period (4 weeks or more)
Electronics
Battery and service indicator (LED)
The lifter has two indicator lights on battery
holder.
The electrical system has a power save function
which will turn off the electrical system after ten
minutes without activity. All lights will turn off.
The system is activated when pushing one of the
operating buttons
Battery indicator (LED)
The battery indicator (LED) will illuminate and
make a sound when battery level is low and the
lifter requires charging. When this occurs the lifter
will have sufficient power available for one full lift-
ing cycle with max load. When battery is critically
low it is only possible to lower the lifting arm.
Service indicator
Number of lifts performed is calculated and
recorded by the lifters electronics.
After a certain period of operation a signal is given
to indicate that service is required.
Service indicator
No light
Green
Yellow
Red
Red + sound
Service scope
Service involves replacing the lifting motor and
inspection/replacement of worn parts. This must
be carried out by authorized personnel.
Service is needed when the service light is red
(calculated 10.000 lifts in SWL.)
11
Molift Smart 150 / www.etac.com
Service indicator
Battery indicator
Mode
Power saving (Stand by)
Ready for use
Order service
Perform service
Perform service
immediately

Publicidad

Tabla de contenido
loading

Tabla de contenido