Power, weight & size:
Lithium-ion battery:
Charging:
Instrument size:
Package size:
Instrument weight:
Gross weight:
Environmental conditions during transport:
Temperature:
Relative humidity:
Pressure:
Environmental conditions during use:
Temperature:
Relative humidity:
Pressure:
IP class:
17) Troubleshooting
No measurement or unexpected value
Re-Used SmartPeg.
The SmartPegs are disposable and should only be used for one or
multiple measurement at one treatment session, for a single patient
only. Repeated re-use may result in false readings due to wear and
tear of the soft aluminium SmartPeg threads.
Wrong SmartPeg type selected for the Implant.
See SmartPeg reference list, osstell.com/smartpegguide
Bone or soft tissue in between SmartPeg and Implant.
Make sure to clean the Implant prosthetic connection before attach-
ing the SmartPeg.
Electromagnetic interference.
Remove the source of electromagnetic interference.
Instrument tip is held too far away from the SmartPeg.
Normally it is sufficient to hold the instrument tip 2-4 mm away from
the SmartPeg, but in some cases as close as 1 mm is necessary.
Instrument does not sense the SmartPeg, hence no
measurement.
Bring the instrument out of the mouth and then in the mouth again.
Try to measure with approx. 45º angle towards the SmartPeg top as
indicated in the upper display.
25087-03 GB
3.7 VDC
Use only Osstell USB cable (USB type C / USB
type A) connected to standard USB 2.0 or 3.0
(type A) port (Max 5,2 VDC).
210 x 35 x 25 mm
272 x 140,2 x 74,60 mm
0,07 kg
0,75 kg
-40 ºC to +70 ºC
10% to 95%
500 hPa to 1060 hPa
+10 ºC to +35 ºC
30% to 75%
700 hPa to 1060 hPa
IP20
13