Instrument does not start
Uncharged battery.
Charge the Osstell Beacon.
Instrument in transport mode.
See section 7 for instruction on how to de-activate transport mode.
Instrument starts up with
Self-tests failed
Contact the local sales representative or distributor for further
instructions.
The instrument does not show red-green-yellow color
during
start-upn
The instrument will flash red-yellow-green during start-up as a func-
tional test of the color indication. If any or all colors are not shown,
the instrument must not be used. Contact the local sales representa-
tive or distributor for further instructions.
Difficult to measure in an exact recommended
No space, e.g. due to adjacent teeth.
Try to measure at a slightly different direction.
Difficulties attaching the SmartPeg
Wrong SmartPeg
Ensure that the SmartPeg is compatible with the implant system.
18) Service and Support
In the event of an instrument malfunction contact the local sales repre-
sentative or distributor for further instructions.
19) Waste and Disposal
The Osstell Beacon instrument should be recycled as electrical equip-
ment. SmartPegs should be recycled as metal. Whenever possible, the
battery should be disposed in a discharged state to avoid heat genera-
tion through inadvertent short-circuiting.
Follow your local and country-specific laws, directives, standards and
guidelines for disposal.