Steps
General Cleaning Instruction
1
Remove and dispose of used barrier sleeve.
2
Soak cloth in Medizime LF enzymatic cleaner. Paying
particular attention to critical sites, wipe down the housing
of the device for at least one (1) minute.
3
Visually inspect Osstell Beacon for contamination and
visible debris. If contamination or visible debris is present,
remove it with a soft bristle brush, cotton swab, or soaked
cloth depending on location of the soil.
4
Soak cloth with distilled water. Pay particular attention to
critical sites. Wipe down device for at least one (1) minute.
5
Inspect device again and repeat steps 2 and 3 if soil persist.
6
Wipe down device with soft cloth dampened with 70% IPA
to help remove moisture.
7
Allow device to air dry completely before next use (mini-
mum three (3) minutes).
Steps
General Disinfection Procedure
1
According to manufacturer's instructions, the minimum
exposure time for the CIDEX® OPA disinfectant is twelve
(12) minutes. Disinfectant application should be performed
by placing the device in a cup with the tip faced down, see
figure below. Fill the cup with CIDEX® OPA to a level which
will allow immersion to 7.5 cm (3 inches), which is the
level where the instrument neck bends, see figure below.
Leave the device immersed for a minimum exposure time
of twelve (12) minutes. Upon completion of the exposure
time, lift the device from the cup with the tip still faced
down and use a clean cloth to manually wipe the device.
2
To remove any residual disinfectant, fill a new cup with
distilled water to a level which will allow immersion to 7.5
cm (3 inches) and leave the device immersed for a mini-
mum of one (1) minute.
3
Repeat the one (1) minute distilled water immersion two
(2) additional times using fresh water for a total of three
(3) rinses.
4
Following removal of the device from water immersion,
thoroughly ensure disinfectant residue removal by wiping
down the device with a soft cloth dampened with 70% IPA.
5
Repeat the 70% IPA wipe procedure two (2) additional
times, for a total of three (3) alcohol wipes.
6
Allow devices to air dry out of exposure to direct sunlight.
Osstell has validated the High Level Disinfection for up to 5000 processing
cycles without damage to the Beacon.