Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Meter view

Press [Meter View]. This key toggles between multiple and single output view.
Multiple Output View
Item
1
Output identifier
When an output is selected, the background becomes highlighted. The selected output is displayed in an enlarged format in
single output view.
2
Output status
OFF: The output is off
CV: The output is in constant voltage mode
CC: The output is in constant current mode
UR: The output is unregulated
3
Output meters
Displays the actual output voltage and current. Displays power in single output view.
4
Output settings
Displays the present output voltage and current settings. Use the numeric keypad or turn the front panel voltage or current
knob to adjust these settings.
5
Interface status
: USB flash drive is connected
: LAN is connected
: Instrument front panel is locked
: Instrument error has occurred
6
Ratings and protection
Displays the present over-voltage protection (OVP) setting and whether over-current protection (OCP) is on or off.
7
Output delay
Displays the OCP, Output On and Output Off delay values.
Keysight EDU36311A User's Guide
Description
OV: Over-voltage protection is tripped
OC: Over-current protection is tripped
OT: Over-temperature protection is tripped
: Instrument in remote mode
: Over-voltage protection is tripped
: Over-current protection is tripped
: Over-temperature protection is tripped
Single Output View