– Setting Change: delay timer start at the end of a settings change in voltage, current, or output state. Press OCP
Delay
Set
| CC.

Clears and OVP and OCP Event

To clear the protection function, first remove that condition that caused the protection fault.
Press Protection Clear to clear the protection function and returns the output to its previous operating state.
From the remote interface:
To set the over-voltage protection for output 1 to the maximum limit:
VOLT:PROT MAX, (@1)
To enable the over-current protection for output 1 and 3:
CURR:PROT:STAT ON, (@1,3)
To set the over-current protection delay time for output 1 at 3 seconds:
CURR:PROT:DEL 3, (@1)
To set the over-current protection delay timer start for output 1 to CC transition:
CURR:PROT:DEL:STAR CCTR, (@1)
To clear protection for output 1:
OUTP:PROT:CLE (@1)
48