Cleaning The Charger, Handpiece, And Eye Shield; Storage Of The Handpiece During Extended Periods Of Non-Use; Disposal; Customer Information - 3M Unitek Ortholux Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 49
Remove the light guide from the solution and rinse
thoroughly (at least 10 sec) in water with low germ
count.
Dry the light guide with a clean cloth.
Check the light guide (see section "Check").
Automatic Cleaning/Disinfection (Disinfector/CDD
(Cleaning and Disinfection Device))
Alternatively, cleaning and disinfecting can be
conducted automatically. Information about validated
procedures can be obtained from 3M Deutschland
GmbH.
Sterilization
Effective cleaning and disinfection are absolutely
essential requirements for effective sterilization.
Only steam sterilization is approved as a sterilization
procedure:
- Maximum sterilization temperature 134 °C
(273 °F)
- Sterilization time (exposure time at sterilization
temperature) at least 20 min. at 121°C (250 °F)
or at least 3 min. at 132 °C (270 °F) /134 °C
(273 °F)
Check
Before using the light guide again, check it for
damaged surfaces, discoloration, and contamination;
do not use damaged light guides. If the light guide
is still contaminated, repeat the cleaning and
disinfection.
Cleaning the Charger, Handpiece, and
Eye Shield
Clean all components with a soft cloth and,
if necessary, a mild cleaning agent (e.g.
dishwashing detergent). Solvents or abrasive
cleaners may not be used in any case, as they
can damage the plastic components!
- Cleaning agents must not enter the unit!
To disinfect all components, spray the disinfectan
on a towel and use it to disinfect the unit. Do not
spray the disinfectant directly on the device.
- Disinfectant agents must not enter the unit!
- Dry residual disinfectants on the device with
a soft and fluff-free cloth, as they damage
the plastic components. If necessary, ask the
manufacturer of the disinfectant if its constant
use will damage plastic surfaces.
Additional Notes: Make sure that charge contact
pins remain dry and are not contacted by metallic
or greasy parts. Do not bend the charge contact
11
English
All manuals and user guides at all-guides.com
pins during drying. Wet charge contact pins will
cause an operating error (malfunction message:
the LED showing the operating status of the
charger flashes red).
Storage of the Handpiece during
Extended Periods of Non-Use

Disposal

Your new unit contains a powerful lithium-ion
rechargeable battery. Dispose of defective batteries
and units in accordance with local legal regulations!

Customer Information

No person is authorized to provide any information
that deviates from the information provided in this
instruction sheet.

Warranty

3M Unitek warrants this product will be free from
defects in material and manufacture for a period
of two years from date of purchase. 3M UNITEK
MAKES NO OTHER WARRANTIES INCLUDING
ANY IMPLIED WARRANTY OF MERCHANTABILITY
OR FITNESS FOR A PARTICULAR PURPOSE.
User is responsible for determining the suitability of
the product for user's application. If this product is
defective within the warranty period, your exclusive
remedy and 3M Unitek's sole obligation shall be repair
or replacement of the 3M Unitek product.

Limitation of Liability

Except where prohibited by law, 3M Unitek will not
be liable for any loss or damage arising from this
product, whether direct, indirect, special, incidental
or consequential, regardless of the theory asserted,
including warranty, contract, negligence or strict
liability.
Information valid as of July 2012
If the handpiece is not to be used for an extended
period of time (e.g., during vacation) fully
charge the battery prior to departure or keep the
handpiece inserted in the operational charger.
A safety switch within the battery prevents a
total discharge.
Discharged or nearly discharged batteries must be
recharged as soon as possible.

Publicidad