H-Style Shoulder Harness
I N S TA L L AT I O N A N D U S E R ' S I N S T R U C T I O N S
These instructions provide important information for the safe use and maintenance of all Bodypoint
H-Style Shoulder Harnesses. Give these instructions to the user or their caregiver and review them
to ensure that they are understood.
WARNING! Product should be installed and fitted
by a qualified rehab technician.
For installation, this product should be mounted with
one or more of the following:
• Flat-Mount
end-fittings
™
• Cinch-Mount
end-fittings
™
• Cam buckles (suitable for users up to 75kg)
Optional Bodypoint belt mounting hardware may also be
used for mounting to wheelchairs and seating systems.
INTENDED USE:
WARNING! Because of the risk of choking, it is
dangerous to use this product without stabilizing the
pelvis – always use with a properly fitted pelvic support
belt.
WARNING! This product should only be used for
positioning a person in a wheelchair. It is NOT intended
for use as a transportation safety device, as a personal
restraint device, or in any other application where its
failure could result in injury. Misuse of this product is
unauthorized and unsafe.
WARNING! This harness must be properly fitted to
support the user's trunk and shoulders without causing
injury. Have your seating specialist demonstrate its
proper adjustment and use.
• A harness that is too tight can restrict respiration and
increase pressure across the shoulders and chest.
• A harness that is too loose can allow the user to slip
down and may create a risk of strangulation.
• Accidental release of this shoulder harness can allow
the user to fall forward.
• A user's inability to self-release can be hazardous if
the user slips down or is trapped in the chair in an
emergency.
If the user's physical or cognitive abilities could prevent
them from safely operating this harness, a caregiver
must be present at all times during its use. Ensure that
all caregivers know how to correctly adjust and unfasten
the product.
558 First Ave. S., Suite 300 | Seattle, WA 98104 | 206.405.4555 | 800.547.5716 | www.bodypoint.com
© 2021 Bodypoint, Inc. All rights reserved.
A R T N O . S H 2 2 4
WARNING! As with any new seating support, this
product may change the way a person sits. Users must
continue to practice regular pressure relief activities and
skin integrity checks, not only where this product
contacts the user, but also in primary pressure-bearing
areas such as the sacrum, legs, and buttocks. If
increased skin redness or irritation occurs, discontinue
use and consult your physician or seating specialist.
Failure to do so may result in serious injury, such as
pressure ulcers.
WARNING! If a serious incident occurs related to the
use of this product, it should be reported to the
manufacturer (Bodypoint, Inc.) and the local Competent
Authority.
PERIODIC SAFETY AND PERFORMANCE CHECKS:
To ensure user safety, this product must be checked pe-
riodically for function and signs of wear. If the product
does not function correctly or if significant wear is found
in the buckles, mounting points, webbing, padding,
or stitches, stop using it and contact your supplier for
qualified repair or replacement by Bodypoint. Under no
circumstance should this product be altered or repaired
by unqualified persons — health and safety depend
on it!
(140°F). Do not bleach. Tumble dry, low temperature,
or drip dry. Do not iron. (Placing the product inside a
cloth bag during washing helps prevent scratches to the
product and the machine.)
SCRAP/DISPOSAL: Product is made of materials which
can be disposed of safely without special precautions at
the end of its useful life.
WARRANTY: This product carries a limited lifetime
warranty against defects in workmanship and materials
arising under normal use by the original consumer. Con-
tact your supplier or Bodypoint for warranty claims.
For more information on Bodypoint products and a list of
distributors outside the USA, go to www.bodypoint.com.
CLEANING: Machine wash, hot, 60°C
Bodypoint Europe BV
Kerkstraat 29
7396 PD Terwolde
The Netherlands
en
Medical
Device