4. Flush the sensor with 50 mL or more particle-free oil to remove any
leftover oil traces.
5. Flush the sensor with the actual oil sample.

Oil to water applications

W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and wear all of
the personal protective equipment appropriate to the chemicals that are handled.
Refer to the current material safety data sheets (MSDS) for safety protocols.
To avoid inaccurate results, refer to the following procedure when
changing the application from oil to water
1. Flush the sensor with mineral spirit several times.
2. Flush the microcell with isopropyl alcohol and then flush the microcell
with clean, particle-free water.
Note: The fluids used one after each other must be miscible. Immiscibility will
create droplets causing false particle counts and may coat the microcell
windows.

Troubleshooting

Error/Warning
Description
Sampler Stalls
Viscosity is too high.
The sensor microcell is
clogged.
10 English
Baseline Hi Error
Message
High noise level, with
clean fluids, but no
sample flow
High noise level, with
flow only
Bad distribution is
exhibited with
Solution
polystyrene spheres.
Dilute the sample with a
clean, particle-free
miscible liquid. Warm the
sample, but do not
exceed the sample flash
point and flammability
specifications.
Refer to
Clean a blocked
microcell
on page 11.
Error/Warning
Description
Sample is too dark. Sample
has a high particle
concentration.
Air bubbles in sample.
Presence of water in an oil
sample or presence of oil in
a water sample.
Sensor microcell is clogged
or dirty.
The "Clean sample" is dirty.
Turbulence/bubbles occur
from debris clogged in the
sensor microcell flow
channel.
Turbulence/bubbles occur
from debris clogged in the
sensor microcell flow
channel.
The sensor microcell is
clogged or dirty.
Solution
Dilute the sample.
Degas the sample.
Discard the sample.
Refer to
Water to oil
application
on page 9 or
to
Water to oil
applications.
Refer to
Clean the sensor
and the microcell
on page 7.
Make sure to run a
filtered blank sample.
Refer to
Clean the sensor
and the microcell
on page 7.
Refer to
Clean the sensor
and the microcell
on page 7.
Refer to
Clean the sensor
and the microcell
on page 7.