Figure 5 Prepare cleaning solution
Figure 6 Fill the microcell with Micro solution
Figure 7 Floss the microcell windows
1 Microcell windows
2 Super Floss
Water to oil application
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and wear all of
the personal protective equipment appropriate to the chemicals that are handled.
Refer to the current material safety data sheets (MSDS) for safety protocols.
To avoid inaccurate results, refer to the following procedure when
changing the application from water to oil.
1. Flush the microcell with clean, particle-free water.
2. Flush the microcell with isopropyl alcohol several times to remove
water.
3. Dry the sensor with clean pressurized air.
3 Flow channel
4 Silver can
English 9