Electromagnetic Emissions; Electromagnetic Immunity - HyperIce Normatec 2.0 Pro Manual De Instrucciones

Ocultar thumbs Ver también para Normatec 2.0 Pro:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 122
equipment. Cables and accessories not specified within the instructions for use are not authorized. Using other cables
and/or accessories may adversely impact safety, performance, and electromagnetic compatibility (increased emission
and decreased immunity). Care should be taken if the equipment is used adjacent to or stacked with other equipment; if
adjacent or stacked use is inevitable, the equipment should be observed to verify normal operation in the configuration in
which it will be used.

ELECTROMAGNETIC EMISSIONS

This equipment is intended for use in the electromagnetic environment specified below. The user of this equipment should
assure that it is used in such an environment.
Emissions
RF emissions
(CISPR 11)
CISPR emissions
classification
Harmonic emissions
(IEC 61000-3-2)
Voltage fluctuations/
flicker
(IEC 61000-3-3)

ELECTROMAGNETIC IMMUNITY

This equipment is intended for use in the electromagnetic environment specified below. The user of this equipment
should assure that it is used in such an environment.
Immunity
IEC 60601-1-2 Test Level
Against
Electrostatic
Contact Discharge:
discharge, ESD
± 8 kV
(IEC 61000-4-2)
Air Discharge:
± 15 kV
Compliance
Electromagnetic Environment
According To
Group 1
The equipment uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
Class B
The equipment is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
Class A
used for domestic purposes.
Complies
Compliance Level
(of this control unit)
± 8 kV
± 15 kV
Electromagnetic Environment
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be kept at levels to reduce
electrostatic charge to suitable levels.