Electrical fast
Power Supply Lines: ± 2 kV
transients/bursts
Input DC Power Ports ± 2 kV
(IEC 61000-4-4)
Signal Input/Output Lines: ± 1 kV
RF Proximity
385 Pulse Modulation, 18 Hz
(IEC 61000-4-3)
450 FM + 5Hz deviation: 1 kHz
sine
710
745
780 Pulse Modulation: 217 Hz
810
870
930 Pulse Modulation: 18 Hz
1720
1845
1970 Pulse Modulation: 217 Hz
2450 Pulse Modulation: 217 Hz
5240
5500
5785 Pulse Modulation: 217 Hz
Surges on AC
Common Mode: ± 2 kV
mains lines
Differential Mode: ± 1 kV
(IEC 61000-4-5)
Power frequency
30 A/m
magnetic field
50/60 Hz
(IEC 61000-4-8)
± 2 kV
Mains power quality should be that
± 2 kV
of a typical commercial or hospital
± 1 kV
environment.
27 V/m
Equipment with high RF emissions
28 V/m
should be kept at a distance to reduce
the likelihood of interference.
9 V/m
28 V/m
28 V/m
28 V/m
9 V/m
± 2 kV
Mains power quality should be that
± 1 kV
of a typical commercial or hospital
environment.
30 A/m
Equipment that emits high levels of
power line magnetic fields (in excess of
3A/m) should be kept at a distance to
reduce the likelihood of interference.