Freestyle® Control Solution; Purpose; Control Solution Information - Freestyle Lite Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

What It Means
Blood drop is too small
2. The test
does not
start after
applying
the blood
Defective test strip
drop.
Sample applied after
meter turns off
Defective meter
23
Contact Customer Care Toll Free at 1-888-522-5226 www.MyFreeStyle.com
ART22813-003_rev-A.indd
23-24
All manuals and user guides at all-guides.com
What To Do
1. Add blood to the same sample area within
60 seconds, or repeat the test using a new
test strip and a larger blood drop.
1. Repeat the test using a new test strip. Wait for
and
to appear before applying blood or
control solution.
2. If problem persists, contact Customer Care.
1. Contact Customer Care.
FreeStyle® Control Solution

Purpose

FreeStyle control solution is a red liquid that contains a fixed amount of glucose.
The purpose of doing a control solution test is:
To practice testing without using your own blood
To confirm that the meter and test strips work together properly
When you are unsure of your blood glucose test results
Contact Customer Care for information on how to obtain control solution.

Control Solution Information

Use only FreeStyle control solution with the meter.
Replace the cap securely on the bottle immediately after use.
Control solution results are accurate only between 59° – 104° F (15° – 40° C).
Do not add water or other liquid to control solution.
There are three levels of control solution available (Low, Normal, High).
For information on how to obtain control solution, contact Customer Care.
IMPORTANT: Control solution results should fall within the control
solution range printed on the test strip vial. Do not use this range
when testing blood glucose.
24
11/9/16 1:29 PM

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido