Out of Range Control Solution Results

Repeat the test if control solution results are outside the range printed on
the test strip vial.
Stop using the meter if control solution results are consistently outside the
range printed on the test strip vial. Contact Customer Care.
Causes of out-of-range results include:
Expired or bad control solution
Expired or bad test strip
Testing error
Watered-down control solution
Meter malfunction
Test strip deterioration
Control solution test done outside 59° – 104° F (15° – 40° C)
25
Contact Customer Care Toll Free at 1-888-522-5226 www.MyFreeStyle.com
ART22813-003_rev-A.indd 25-26
All manuals and user guides at all-guides.com

Performing a Control Solution Test

1.
Check control solution expiration date.
IMPORTANT: Do not use control solution past expiration date. Discard
control solution 3 months after opening or on the expiration date
printed on the bottle whichever comes first. (Example: open April 15,
discard July 15; write the discard date on the side of the bottle.)
2.
Remove test strip from vial.
3.
Insert test strip to turn meter on.
Note: The meter turns off after 2 minutes of inactivity.
Remove and reinsert the unused test strip to restart
the meter.
If the meter does not turn on go to Troubleshooting section.
Top
Printed side up
Sample areas
26