IMPORTANT: If you require assistance or if you notice any signs of deterioration on
the meter or lancing device (such as clouding or crazing on the display of the meter,
corroding or eroding of the plastic housing, or cracking of plastic housing or display) or
if the meter does not turn on or if the lancing device does not work after cleaning and
disinfection, discontinue use of the meter or lancing device and contact Customer Care
(1-888-522-5226).
Cleaning
Step 1
For cleaning, wipe all outside surfaces of the meter and lancing device thoroughly with
the bleach wipes until they are visibly clean. Clean the strip port, data port and battery
compartment with caution and do not force liquids into sensitive meter electronic
openings. Clean your meter at least once a week. Clean your lancing device when visibly
dirty (at least once a week).
Step 2
Allow the meter and lancing device to air dry completely before performing a blood
glucose test.
Step 3
When finished, thoroughly wash your hands with soap and water.
37
Contact Customer Care Toll Free at 1-888-522-5226 www.MyFreeStyle.com
ART22813-003_rev-A.indd 37-38
All manuals and user guides at all-guides.com
Disinfection
Step 1
For disinfection, wipe all outside surfaces of the meter and lancing device thoroughly
with the bleach wipe until they are wet in order to disinfect the meter and lancing
device. Clean the strip port, data port and battery compartment with caution and do not
force liquids into sensitive meter electronic openings.
Step 2
Allow the meter and lancing device surfaces to remain wet for 60 seconds (1 full minute).
After 1 full minute, immediately dry with drying towel to remove any residual moisture.
Disinfect your meter at least once a week. Disinfect your lancing device when visibly
dirty (at least once a week).
Step 3
Allow the meter and lancing device to air dry completely before performing a blood
glucose test.
Step 4
When finished, thoroughly wash your hands with soap and water.
38
11/9/16
1:29 PM