4.
Confirm System Check Screen.
This display appears when you turn on the meter. Do not use the meter if
the display check screen does not exactly match the example. Contact
Customer Care.
See The FreeStyle Lite® Meter and How It Works section for more information.
If the system check screen does not appear, go to Troubleshooting section.
5.
Apply and hold control solution to 1 sample area of the test strip until:
You see short lines moving clockwise on the display or
You hear a beep
This indicates the test strip obtained enough control solution and the meter is
processing. If the test does not start after applying the control solution, go to
Troubleshooting section.
IMPORTANT: If a short line does not appear after 5 seconds, the
sample may be too small. Do not add control solution to both sides
of the strip. The test strips are for single use only. Do not reuse test
strips. Discard used test strips.
27
Contact Customer Care Toll Free at 1-888-522-5226 www.MyFreeStyle.com
ART22813-003_rev-A.indd 27-28
All manuals and user guides at all-guides.com
6.
View result.
The result appears on the display when the test is complete.
(Example only)
7.
Compare the control solution result to the range printed on the test strip vial.
The result should fall within the range.
Note: If you see an "Er1, Er2, Er3, or Er4" on the display, it is an error code. Go to
Error Codes section.
8.
c
Press and hold
for 2 seconds until
result as a control solution result. To unmark a control solution test result, press
c
and hold
for 2 seconds. A value with the
this reading as a control solution test.
(Example only)
9.
Remove the test strip to turn meter off.
appears on the display to mark the
indicates you have marked
28