5 Descripción detallada
5.11.1.7
5.11.1.8
5.11.1.9
5.11.1.10
5.11.1.11
5.11.1.12
114
Mode.SampleSeq.1.∗
etc. up to .100
Indexed sample sequence; its commands will be executed line by line in
processing. See start sequence (5.11.1.5).
Mode.FinalSeq.1.Cmd
etc. up to .100
Defines the command of the indexed command line in a final sequence.
See start sequence (5.11.1.4).
Mode.FinalSeq.1.∗
etc. up to .100
Indexed final sequence; its commands will be executed line by line in proc-
essing. See start sequence (5.11.1.5).
Mode.Report.Assign
Mode.Report.TDelta
Report definition.
Only one report is possible. In the case of a 'full' result report, a measuring
point list with the oven temperatures as a function of time is output by con-
trast with the 'short' result report. The time interval (in seconds) between the
measuring points defines the node .TDelta. 400 measuring points are pos-
sible, see 5.11.3.2.
Mode.Changer.RackNo
Mode.Changer.L1Rate
Mode.Changer.L2Rate
Mode.Changer.ShRate
Mode.Changer.ShDir
Mode.Changer.ModeSample
Changer settings.
RackNo:
Rack number, forces the use of the defined rack with the
current method (0 = any rack).
L1Rate:
Lift speed for tower 1, in mm/sec
L2Rate:
Lift speed for tower 2, in mm/sec
ShRate:
Turning speed of the rack, in angular degrees/sec
ShDir:
Turning direction of the rack (ascending or descending
rack positions; auto. means automatic choice of the
shortest path)
ModeSample:
Reaction on missing sample beaker. (MOVE = next sam-
ple beaker will be chosen regarding the recent SAMPLE
command, display = display warning.)
Mode.TimeoutSet.HTime
Mode.TimeoutSet.HAction
Mode.TimeoutSet.STime
Mode.TimeoutSet.SAction
Timeout settings. Timeout times are defined in seconds.
The HEATER timeout occurs if the set oven temperature (in the case of the
HEATER command) is not reached after the specified heating (in minutes).
HTime is the additionally granted waiting time for still reaching the target
.Move..., .Lift..., .Sample...,
.Heater...,.Def..., .Flow..., Dos...,
Scan..., .Ctrl..., .Wait..., .Rack, .Nop
NOP, MOVE, LIFT, STIR,
DEF, PUMP, DOS, SCAN,
SAMPLE, CTRL, WAIT, ENDSEQ
.Move..., .Lift..., .Sample...,
.Heater...,.Def..., .Flow..., Dos...,
Scan..., .Ctrl..., .Wait..., .Rack, .Nop
none, short, full, param,
config, Usermeth
MOVE, display
774 Oven Sample Processor, Instrucciones para el uso
1...10...60
0...16
3...12 mm/s
3...12 mm/s
3...20 w/s
+, –, auto.
off, 0...999
cont, STOP
off, 0...999
cont, ERROR