Definition of new rack types:
5.11.2.9
5.11.2.10
774 Oven Sample Processor, Instrucciones para el uso
5.11 Description of the remote control commands
Config.RackDef.SpezBeak.1.Pos
etc. until .8
Rack positions of special beakers 1 to 8 (position 0 = not defined).
Config.PosTab.TabIdx
Index of position tables..
Config.PosTab.Name
Config.PosTab.R1Num
Config.PosTab.R2Num
Config.PosTab.R3Num
Config.PosTab.R1Off
Config.PosTab.R2Off
Config.PosTab.Num
Definitions of position tables. Depending on the table index the correspond-
ing one of the 16 possible data sets is overlayed.
Name:
Identification of the rack type. Will be available as a selec-
tor under &Config.RackDef.Type (see 5.11.2.4).
R1Num:
Highest Beaker position in row 1
R2Num:
Highest Beaker position in row 2
R3Num:
Highest Beaker position in row 3
R1Off
Offset in
row 1 (for the beaker test)
R2Off
Offset in
row 2 (for the beaker test)
Num:
Number of rack positions
Config.PosTab.1.Value
bis .200.Value
Angular offset for the particular rack position in tenth of angular degrees
(
1
/
-degrees).
10
The definition of rack types is only possible via RS232 interface.
• Set table index (see 5.11.2.6).
• Enter name of rack type (&Config.PosTab.Name, see 5.11.2.7).
• Enter highest beaker position in row 1 to 3 (&Config.PosTab.R1–3Num,
see 5.11.2.7).
• Enter the offset angle between beaker position 1 in row 1 and beaker
position (R1num+1) in row 2, respectively and tower 1
(&Config.PosTab.R1–2Off, see 5.11.2.7).
• Define number of rack positions (&Config.Pos.Tab.Num, see 5.11.2.7).
• Enter the particular angular offset for each rack position
(Irregular layouts of rack positions are possible)
Config.WetPart.WetPartNo
Identification of dosing unit.
Config.WetPart.1.MaxRate
Config.WetPart.1.Length
Config.WetPart.1.Diameter
until .4.Diameter
(Default value .2.Length : 250 mm)
0...number of rack positions
(R1Num + 2)...(R3Num – 2)
1
/
-angular degrees for the beaker positions in
10
1
/
-angular degrees for the beaker positions in
10
0...31
8 ASCII characters
2...(R2Num – 2)
(R2Num + 2)...200
0...3599
0...3599
1...200
0...3599
1...12
0.01...160 ml/min
0...1000...30000 mm
0.1...2...20 mm