Unpack the cartons and check all items. Make sure that the product operates before you place into service.
W W A A R R N N I I N N G G
• Consult the operations manual for the support surface before you attach the pump.
• Do not use in the presence of flammable anesthetics, nitrous oxide, or oxygen-rich environments. Risk of explosion can
result.
• Do not expose this product to liquids while powered as this could result in a severe electrical hazard.
• Always setup and use in accordance with the instructions to avoid harmful interference to other devices in the vicinity.
This product radiates radio frequency energy.
C C A A U U T T I I O O N N
• Always allow the product to come to room temperature after exposure to extreme high or low temperatures.
• Always check the power cord for cuts, exposed wires, worn insulation, or other damage that could result in electrical
hazards.
1. Bring the product to room temperature.
2. Place the pump in a secure location. Allow space around the pump to connect and disconnect the hoses and power
cord. Optional placements:
a. Option 1: Hang the pump on a compatible bed. Make sure that the pump is secure.
b. Option 2: Place the pump on the floor. Make sure that the pump is away from foot traffic to avoid the risk of a trip
hazard. Do not place the pump in the path of lifting or lower mechanisms of the bed.
C C o o n n n n e e c c t t i i n n g g t t h h e e p p o o w w e e r r c c o o r r d d
1. Attach the power cord to the pump.
2. Plug the power cord into an accessible hospital grade receptacle.
C C o o n n n n e e c c t t i i n n g g t t h h e e h h o o s s e e
W W A A R R N N I I N N G G - - Always make sure the hose is routed down and away from the patient. Avoid accidental patient contact at the
foot end of the support surface. Hose temperature may reach 48° C (118.4° F).
1. Attach the hose to the pump hose ports.
2. Open the hose port (A) on the support surface (Figure 2).
3. Attach the hose to the support surface hose port. Make sure that rib (B) lines up with the red line on the hose (C).
2874-009-005 Rev A.0
B
C
A
F F i i g g u u r r e e 2 2 – – H H o o s s e e c c o o n n n n e e c c t t i i o o n n
9