Stryker IsoTour Manual De Operación página 22

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 157
P P r r e e v v e e n n t t i i v v e e m m a a i i n n t t e e n n a a n n c c e e
W W A A R R N N I I N N G G - - Do not modify or change this device. Service should only be completed by qualified personnel. Failure could
result in injury and void your warranty.
N N o o t t e e
• At a minimum, check all items listed during annual preventive maintenance for all Stryker Medical products. You may
need to perform preventive maintenance checks more often based on your level of product usage.
• Remove product from service before you perform preventive maintenance.
• Consult your local regulations to dispose of electronic equipment.
Inspect the following items:
All fasteners are secure
Pump housing or components (hoses, power cords, or case) for cracks, holes, or damaged
Hooks that hang the pump on the bed frame are not damaged
No air leaks from the pump or the attached connectors or hoses
Graphical user interface is not cracked or damaged
HEPA filter
Fuse
All functions on graphical user interface operate
Run diagnostic test (All okay)
Product serial number:
Completed by:
Date:
EN
18
2874-009-005 Rev A.0

Publicidad

Tabla de contenido
loading

Productos relacionados para Stryker IsoTour

Este manual también es adecuado para:

2874

Tabla de contenido