4. Insert the Adjustment Tool into the
Adjustable Height Locator Tool so that the
arrow is pointing to the current performance
setting. With one hand holding the "petals"
of the Adjustable Height Locator Tool, turn
the Adjustment Tool with the other hand until
the arrow points to the desired performance
setting (Figure 4). The Adjustment Tool will
provide an audible click and a tactile response
as you turn to each setting.
Note: The Adjustment Tool is equipped with
a mechanical stop between Settings 1 and 8.
This is to prevent inadvertent adjustment
between the extremes of the available settings.
Note: Successful adjustment is accomplished
by starting at the valve's current setting and
moving directly to the desired setting.
5. Remove the Adjustment Tool from the
Adjustable Height Locator Tool by lifting it
straight upwards a minimum of 3 cm (1.25 in.)
before moving it horizontally away from the
Adjustable Height Locator Tool in order to
avoid inadvertently changing the valve setting.
6. Repeat Steps 2 and 3 to confirm successful
adjustment of the performance setting.
Always confirm the desired performance
setting of the valve.
7. If the desired performance setting is not
achieved, repeat Steps 4 through 6.
8. It is recommended to record the valve setting in
the patient's record and I.D. wallet card (available
from local Integra sales representative).
9. Return all tools to their proper locations in the
storage case to prevent damage.
Note: fully collapse the Adjustable Height
Locator Tool (see figure 1B).
7