Descargar Imprimir esta página

Agilent Technologies TwisTorr Medium-TMP Manual De Instrucciones página 248

Ocultar thumbs Ver también para TwisTorr Medium-TMP:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 51
17
Technical Information
How to Use by Front Panel
Pump Start, Stop and Reset Fail
Press once the START/STOP button for the START command and
again for the STOP command. In the case of controller's failure, press
once to reset the fail condition.
The front panel START/STOP function is available only if the controller is set to
NOTE
"FRONT mode". If it is set to "REMOTE mode", when the key is pressed an error
message is shown indicating "COMMAND LOCKED IN REMOTE", if it is set to
"SERIAL" mode, the "COMMAND LOCKED IN SERIAL MODE" error message is
displayed.The error message is displayed for approx. 1 second before moving to
the previous parameter. The "Reset Fail" function is always available regardless
of controller mode.
248/284
TwisTorr Medium-TMP Rack Controller User Manual / 87-901-049-01

Publicidad

loading

Este manual también es adecuado para:

X3501-64016X3501-64006